Table 2: Validation data for determination of lamivudine by the proposed HPLC procedure.

Item

LVD

λ max (nm)

280

Concentration range (µg/ml)

5-150

Regression equation:

 

Intercept(a)

1752.70

Slope (b)

19189.49

Variance of intercept (Sa2)

3853.46

Variance around slope (Sb2)

50.08

Correlation coefficient(r)

0.99999

Variance (Sy/x)

6495.06

Accuracy (mean ± SD %)

98.99 ± 0.74%

Precision (RSD %)

0.26%

Limit of detection (µg/ml)

1.09

Limit of quantitation (µg/ml)

3.6